GeneCE-100 Bio-Fragment Analyzing System
GeneCE-100 provides flexible and reliable nucleic acid quality control for DNA QC, NGS libraries, cfDNA, PCR fragments and RNA QC with high sensitivity down to 1 pg/µL and 1–96 sample throughput.
Send us an inquiry and our team will respond within 24 hours.
GeneCE-100 Bio-Fragment Analyzing System
The GeneCE-100 Bio-Fragment Analyzing System delivers comprehensive nucleic acid quality control across a wide range of applications, including DNA QC, NGS library characterization, cell-free DNA (cfDNA) analysis, PCR fragment verification, and RNA integrity assessment. By leveraging a single-channel cartridge platform, the instrument ensures consistent, high-precision results while accommodating flexible sample throughput from 1 to 96 samples per run.
Whether you are preparing libraries for next-generation sequencing or verifying fragment sizes after enzymatic digestion, GeneCE-100 provides the resolution and sensitivity needed to make confident decisions at every stage of your workflow.
Key Features
10-Inch Rotatable Touch Screen — The integrated 10-inch high-resolution touch display can switch between horizontal and vertical orientations to suit your workflow preferences. The intuitive interface allows well layout, parameter configuration, and data analysis directly on the instrument without requiring an external computer.
Exceptional Sensitivity and Resolution — With a detection limit as low as 1 pg/µL (under optimized conditions using the NH1 cartridge) and fragment resolution down to 1–4 bp, GeneCE-100 enables precise sizing and quantification even for closely spaced fragments. The high-sensitivity PMT detector and LED light source work in concert to deliver reliable data across diverse sample types.
Intelligent Cartridge Management — Each gel cartridge embeds an RFID chip that automatically records the number of remaining tests and expiration dates, providing real-time prompts to prevent expired reagent usage and streamline inventory management.
Parallel Sample Processing — The flexible 1–96 sample capacity allows both small-scale validation runs and full 96-well plate processing, making GeneCE-100 equally suitable for routine QC and higher-throughput production environments.
Software Capabilities
The built-in software running on Windows 11 provides a powerful, self-contained analysis environment. Users can perform sample well layout, configure electrophoresis parameters, and conduct data analysis entirely through the 10-inch touch screen. The graphical interface presents results including fragment concentration, size distribution, 28S/18S ratios, DNA Integrity Indicator (DII), RNA Integrity Indicator (RII), and DV200 values — enabling rapid, informed downstream decisions.
Technical Specifications
| Parameter | GeneCE-100 |
|---|---|
| Detection Method | Fluorescence |
| System Light Source | LED |
| Detector | High-sensitivity PMT |
| Throughput | 1–96 samples |
| Detection Limit | 1 pg/µL (optimized condition) |
| Resolution | 1–4 bp |
| Analysis Time | 3–5 min/sample |
| Sample Consumption | <0.2 µL |
| Fragment Size CV% | ≤2% |
| Fragment Concentration CV% | ≤4% |
| Instrument Type | Single channel |
| Operating System | Windows 11 and higher |
| Display | 10-inch high-resolution touch screen |
| Power Supply | 110–240 V/AC, 50/60 Hz |
| Weight | 26 kg |
| Dimensions | 460 × 445 × 448 mm |
Reagent Cartridges
GeneCE-100 supports three cartridge types to match different analytical needs. All cartridges operate across the 20–5,000 bp separation range with sample consumption below 0.2 µL per analysis.
| Cartridge | Separation Range | Max. Resolution | Sample Size / Cartridge | Detection Limit |
|---|---|---|---|---|
| SC2 Standard Cartridge Kit | 20–5,000 bp | 5–10 bp | 200 runs | 5 pg/µL |
| SH1 High-Resolution Cartridge Kit | 20–5,000 bp | 1–4 bp | 200 runs | 5 pg/µL |
| NH1 High-Sensitivity Cartridge Kit | 20–5,000 bp | 1–4 bp | 100 runs | 1 pg/µL |
Ordering Information
| Code | Description |
|---|---|
| 63-RS-31010-00 | GeneCE-100 Bio-Fragment Analyzing System |
| 63-RS-32010-00 | Standard Cartridge Kit (SC2), 200 runs |
| 63-RS-32020-00 | High-Resolution Analysis Cartridge Kit (SH1), 200 runs |
| 63-RS-32030-00 | High-Sensitivity Analysis Cartridge Kit (NH1), 100 runs |